xsh_lingain7
Computes detector's gain/linearity-map
Usage
esorex xsh_lingain [OPTIONS] FILE.sof
Overview
This recipe computes detector linearity coefficients and gain Input Frames :
arm = UVB-VIS:
- A set of n RAW linearity frames (Format=RAW, n >=8, Tag = LINEARITY_arm_ON)
- A set of n RAW bias frames (Format=RAW, n >=1, Tag = LINEARITY_arm_OFF)
arm = NIR:
- A set of n RAW linearity on frames (Format=RAW, n >=8, Tag = LINEARITY_arm_ON)
- A set of n RAW linearity off frames (Format=RAW, n >=8, Tag = LINEARITY_arm_OFF)
- [OPTIONAL] a static bad pixel map (PRO.CATG=BP_MAP_RP_arm)
Note: on and off frames are taken in pairs sequence: OFF-ON-ON-OFF,
at least 16 frames.
Products :
- A linearity map table, PRO.CATG = BP_MAP_LIN_NIR
- A gain table, PRO.CATG = GAIN_INFO
- A cube cointaining the linearity coefficients, PRO.CATG = COEFFS_CUBE_arm
- A linearity map image (RAW format), PRO.CATG = RAW_BP_MAP_NL_arm
- A linearity map image (PRE format), PRO.CATG = BP_MAP_NL_arm
Options
-
--keep-temp <str> -
If ´no´, temporary files are deleted. (str; default: ´no´). The full name of this option for the EsoRex configuration file is
xsh.xsh_lingain.keep-temp[default = no].
-
--debug-level <str> -
Additional xshooter debug level. One of ´none´, ´low´, ´medium´, ´high´ (str; default: ´none´). The full name of this option for the EsoRex configuration file is
xsh.xsh_lingain.debug-level[default = none].
-
--time-stamp <bool> -
Add timestamp to product file name. (bool; default: False). The full name of this option for the EsoRex configuration file is
xsh.xsh_lingain.time-stamp[default = False].
-
--decode-bp <int> -
Integer representation of the bits to be considered bad when decoding the bad pixel mask pixel values. Most frequent codes relevant for the user: 0: good pixel, 8: pick-up noise, 16: cosmic-ray removed, 32: cosmic-ray unremoved, 128: calibration file defect, 256: hot pixel, 512: dark pixel, 4096: A/D converted saturation, 32768: non linear pixel, 1048576: extrapolated flux in NIR, 4194304: Interpolated flux during extraction. (int; default: 2144337919). The full name of this option for the EsoRex configuration file is
xsh.xsh_lingain.decode-bp[default = 2144337919].
-
--method <str> -
Method to be used when computing GAIN. Methods appliable: <PTC | MED>. By default PTC method will be applied. (str; default: ´PTC´). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.method[default = PTC].
-
--order <int> -
Polynomial order for the fit (Linearity) (int; default: 3). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.order[default = 3].
-
--kappa <float> -
Kappa value for the kappa-sigma clipping (Gain) (float; default: 3.0). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.kappa[default = 3.0].
-
--niter <int> -
Number of iterations to compute rms (Gain) (int; default: 5). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.niter[default = 5].
-
--llx <int> -
x coordinate of the lower-left point of the region of interest. If not modified, default value will be 1. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.llx[default = -1].
-
--lly <int> -
y coordinate of the lower-left point of the region of interest. If not modified, default value will be 1. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.lly[default = -1].
-
--urx <int> -
x coordinate of the upper-right point of the region of interest. If not modified, default value will be X dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.urx[default = -1].
-
--ury <int> -
y coordinate of the upper-right point of the region of interest. If not modified, default value will be Y dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.ury[default = -1].
-
--ref_level <int> -
User reference level (int; default: 10000). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.ref_level[default = 10000].
-
--intermediate <bool> -
De-/Activate intermediate products (bool; default: False). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.intermediate[default = False].
-
--autocorr <bool> -
De-/Activate the autocorr option (bool; default: False). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.autocorr[default = False].
-
--collapse <bool> -
De-/Activate the collapse option (bool; default: True). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.collapse[default = True].
-
--rescale <bool> -
De-/Activate the image rescale option (bool; default: True). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.rescale[default = True].
-
--pix2pix <bool> -
De-/Activate the computation with pixel to pixel accuracy (bool; default: True). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.pix2pix[default = True].
-
--bpmbin <bool> -
De-/Activate the binary bpm option (bool; default: False). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.bpmbin[default = False].
-
--m <int> -
Maximum x-shift for the autocorr (int; default: 26). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.m[default = 26].
-
--filter <int> -
Upper limit of Median flux to be filtered (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.filter[default = -1].
-
--n <int> -
Maximum y-shift for the autocorr (int; default: 26). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.n[default = 26].
-
--tolerance <float> -
Tolerance for pair discrimination (float; default: 0.1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.tolerance[default = 0.1].
-
--pafgen <bool> -
Generate PAF file (bool; default: False). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.pafgen[default = False].
-
--pafname <str> -
Specific name for PAF file (str; default: ´xsh_lingain´). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.pafname[default = xsh_lingain].
-
--exts <int> -
Activate the multi-exts option. Choose -1 to process all extensions. Choose an extension number to process the appropriate extension. (int; default: 0). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.exts[default = 0].
-
--fpn_method <str> -
Method for computing Fixed Pattern Noise (SMOOTH or HISTOGRAM) (str; default: ´HISTOGRAM´). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.fpn_method[default = HISTOGRAM].
-
--fpn_smooth <int> -
template size in pixels for smoothing during FPN computation (only for SMOOTH method) (int; default: 13). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.fpn_smooth[default = 13].
-
--saturation_limit <float> -
all frames with mean saturation above the limit would not be used in linearity calculation (float; default: 65535.0). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.saturation_limit[default = 65535.0].
-
--gain_threshold <float> -
all frames with mean flux above the threshold would not be used in gain calculation (float; default: 65535.0). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.gain_threshold[default = 65535.0].
-
--coeffs_cube_split <bool> -
if TRUE, the recipe writes as many COEFFS_CUBE_Pi (i=0..order) as the value of the order parameter in a separate file (bool; default: False). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.coeffs_cube_split[default = False].
-
--llx1 <int> -
x coord of the lower-left point of the first field used for contamination measurement. If not modified, default value will be 1. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.llx1[default = -1].
-
--lly1 <int> -
y coord of the lower-left point of the first field used for contamination measurement. If not modified, default value will be 1. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.lly1[default = -1].
-
--urx1 <int> -
x coord of the upper-right point of the first field used for contamination measurement. If not modified, default value will be X dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.urx1[default = -1].
-
--ury1 <int> -
y coord of the upper-right point of the first field used for contamination measurement. If not modified, default value will be Y dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.ury1[default = -1].
-
--llx2 <int> -
x coord of the lower-left point of the second field used for contamination measurement. If not modified, default value will be 1. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.llx2[default = -1].
-
--lly2 <int> -
y coord of the lower-left point of the second field used for contamination measurement. If not modified, default value will be 1. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.lly2[default = -1].
-
--urx2 <int> -
x coord of the upper-right point of the second field used for contamination measurement. If not modified, default value will be half of the X dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.urx2[default = -1].
-
--ury2 <int> -
y coord of the upper-right point of the second field used for contamination measurement. If not modified, default value will be half of the Y dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.ury2[default = -1].
-
--llx3 <int> -
x coord of the lower-left point of the third field used for contamination measurement. If not modified, default value will be 1. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.llx3[default = -1].
-
--lly3 <int> -
y coord of the lower-left point of the third field used for contamination measurement. If not modified, default value will be half of the Y dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.lly3[default = -1].
-
--urx3 <int> -
x coord of the upper-right point of the third field used for contamination measurement. If not modified, default value will be half of X dimension of the image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.urx3[default = -1].
-
--ury3 <int> -
y coord of the upper-right point of the third field used for contamination measurement. If not modified, default value will be Y dimension of the image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.ury3[default = -1].
-
--llx4 <int> -
x coord of the lower-left point of the fourth field used for contamination measurement. If not modified, default value will be half of X dimension of the image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.llx4[default = -1].
-
--lly4 <int> -
y coord of the lower-left point of the fourth field used for contamination measurement. If not modified, default value will be half of the Y dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.lly4[default = -1].
-
--urx4 <int> -
x coord of the upper-right point of the fourth field used for contamination measurement. If not modified, default value will be X dimension of the image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.urx4[default = -1].
-
--ury4 <int> -
y coord of the upper-right point of the fourth field used for contamination measurement. If not modified, default value will be Y dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.ury4[default = -1].
-
--llx5 <int> -
x coord of the lower-left point of the fifth field used for contamination measurement. If not modified, default value will be half of the X dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.llx5[default = -1].
-
--lly5 <int> -
y coord of the lower-left point of the fifth field used for contamination measurement. If not modified, default value will be 1. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.lly5[default = -1].
-
--urx5 <int> -
x coord of the upper-right point of the fifth field used for contamination measurement. If not modified, default value will be X dimension of the image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.urx5[default = -1].
-
--ury5 <int> -
y coord of the upper-right point of the fifth field used for contamination measurement. If not modified, default value will be half of Y dimension of the input image. (int; default: -1). The full name of this option for the EsoRex configuration file is
xshoo.xsh_lingain.ury5[default = -1].
Note that it is possible to create a configuration file containing these options, along with suitable default values. Please refer to the details provided by the 'esorex --help' command.
Version
xsh_lingain 3.5.0
Releases
4 releases since Debian 8 (Jessie); the text changed 3 times.
--gain_threshold since Debian 9 (Stretch).